Abstract
In this work we present the application of a 2D single grating wavefront sensor to align and characterize the 100 nm focus at the Coherent X-ray Imaging (CXI) endstation at the Linac Coherent Light Source (LCLS). The results agree well with a model of the system, indicating that the mirrors perform as designed when alignment is optimized. In addition, a comparison with the imprint technique confirms the validity of the results, which showed that wavefront-based alignment resulted in negligible astigmatism. Analysis of the retrieved focus profile indicates that intensities <10²¹ W=cm² are achievable with currently available LCLS beam parameters and optimal mirror alignment.