Radiation detectors based onμchannel plates (MCP) are planned for installation at the European XFEL. Main purpose of these detectors is searching a signature of lasing and further fine tuning of the FEL process. Detectors operate in a wide dynamic range from the level of spontaneous emission to the saturation level (between a few nJ and 25 mJ), and in a wide wavelength range from 0.05 nm to 0.4 nm for SASE1 and SASE2, and from 0.4 nm to 4.43 nm for SASE3. The SR tests validation of the MCP-based detector applied for XFEL lines SASE1 and SASE2 were performed at the DORIS beamline BW1 at SR with photon energy of 8.5-12.4 keV. The absolute measurements of a photon pulse energy for hard X-ray radiation were performed with application of MCP and photodiode detectors. Pulse-to-pulse photon energy measurements with MCPs and silicon photo detector were done with 192 ns and 96 ns repetition intervals. The SR beam imaging measurement at X-ray irradiation was performed at test validation experiments.