European XFEL (EuXFEL) generates high-intensity ultra-short pulses at MHz repetition rate. At hard X-ray instruments, cryo-cooled silicon monochromators are used to reduce pulse bandwidth. Here, first experimental observations during commissioning of a cryo-cooled monochromator at Materials Imaging and Dynamics (MID) instrument are presented and compared with heat flow simulations. A thermal relaxation time is estimated and compared with arrival time interval between pulses. This provides the repetition rate tolerable for stable operation of monochromator.